Recommended Fittings for 25 μm ID Tubing

by IDEX Health & Science
Tube ODPortPort MaterialNutFerruleSleeveTorquePSI Rating
.062”10-32 conedSSF-196F-192N/A3 inch / pounds7000
.062”10-32 conedPEEKF-196F-192N/A4 inch / pounds6500
.032”10-32 conedPEEKF-331NSTDF-2474 inch / pounds4000
.032”6-32 conedPEEKF-126HN/AN/A2 inch / pounds7000
360 μm10-32 conedPEEKF-331NSTDF-2424 inch / pounds4000
360 μm6-32 conedPEEKF-124HN/AN/A2 inch / pounds7000

 

Instructions for Use

  1. 1/16” OD tubing should be scored and snapped with the IDEX Health & Science tubing cutter (A-327). Twirl cutting may cause occlusion of 25 μm ID.
  2. 1/32” and 360 μm tubing should be cut with the IDEX Health & Science tubing cutter (A-350)
  3. Tightening more than published torque may occlude tubing ID.